|
|
Contents OverviewIn addition to the built in current meter rating curves for the default Price AA and Pygmy meters (Note: Existing firmware has the rating curves for the Standard Rating Table / Equation No. 1), the AquaCalc 5000, using the Basic (G) software, can store the rating curves for a user-defined / “nonstandard” current meter in each transect. The hydrographer must re-enter the meter constants in each of the transects that he/she wishes to use a user-defined or non-standard meter. To use a user-defined / nonstandard current meter, the user needs to know the “rating curve” or the equation that defines the relationship between the number of revolutions per second (R) that the cups spin, and the measured velocity. If you are using a metric meter please see Non-Standard Metric heading of the Using Your AquaCalc chapter. Meter Rating EquationsAs used by the USGS, meter rating curves are either a single or a pair of equations that define the relationship between the number of revolutions per second of the current meter cups and measured water velocity. The equation for a single or for each segment of a pair of equations equation takes the form: Y = m*X + b or Velocity=(m*X*( Revolutions/Second)) + b where m is the slope of the line, and b is the intercept. Some current meters, such as a Type AA, use two equations: one of which is used for lower velocities, the other for higher. A crossover velocity is also specified. An example of a Type AA Rating Table with the equations is shown below. Entering Single Segment Rating Equations
|
Nonstandard current meter rating curves are stored in the AquaCalc 5000 using the Basic (G) firmware as a series of four “constants” that represent the two equations or intersecting lines of a meter rating curve. A “hidden” fifth constant (C5) is fixed with the value of 2.2. This is to be used with ice meters only.
Constants C1 and C2 correspond to m1, and b1 for the first equation respectively and C3 and C4 to m2, and b2 for the second equation. The constant C5 is the velocity value at the intersection of the two straight lines (currently fixed at 2.2)
If you are entering a single line segment (as in the new Standard Rating Table /Equation No. 2) Enter the same constants for C3 and C4 as you entered for C1 and C2 (C3=C1 and C3=C2).
So for the sample Type AA equation used above:
EQUATIONS: V=2.18R + .020(2.200) 2.170R + .030
The constants to be entered are:
Nonstandard meter settings are setup in the Transect Information Settings menu.
Important: You must perform these steps in every transect you wish to use a user-defined or non-standard meter (including the new Standard Rating No. 2 equation.) Meter constants are not shared between transects.
- Go to the Transect that you wish to measure in while using a user-defined or non-standard meter (including the new Standard Rating No. 2 equation)
- Press the Setup (0) key to access the Main Menu.
- Press the 1 (2 = TrInfo) key to enter the Transect Information Setup.
- Press Enter until the following screen appears (The bottom line of this screen will list your currently chosen meter type):
|
Press the +/- key to cycle through the different meter types until you reach the Non-Standard selection:
|
- Press the Enter key to view the first meter constant.
|
- Pressing the +/- key will allow you to enter the value for the constant.
Existing AquaCalc firmware can only handle three places after the decimal, you must round off to three decimal places. Once you have entered the value, press Enter.
The AquaCalc will prompt you for each of the remaining constants C2 through C4 (C5 is fixed at a value of 2.2). Enter them as you entered Constant 1 (C1) above.
If you are entering a current meter rating that has only one segment (such as the new Standard Meter Rating Table No. 2 equations), repeat the values for C1 and C2 in C3 and C4 (so that C3=C1 and C4=C2). See the sections above for a description of single segment meter constants.
After entering all five “constants”, you will be presented with the Set Measurement Time Screen. Enter the measurement time or press the Setup key (0) to return to the Main Menu.
The following table shows the values to be entered for Price type AA and pygmy meters and for the new Standard Rating No. 2 equations and the old Standard Rating No. 1.
Note that you must round the new Standard No. 2 values to three decimal places to accommodate the limitations of the current version of AquaCalc 5000 firmware. An update of the firmware is necessary to allow for four decimal place entries. This results in an error of approximately 0.02 percent overestimation for the velocity values. Please see our Customer Support page discussing the New Standards for more information.
|
Setting the Meter Identifier - METER ID#
You may also set a meter identifier for he meter you use in Meter Identifier screen. This screen appears before the Change Meter Type screen described above and allows you to record the Serial Number or some other identifying number of the current meter used to measure the stream. Up to six digits including a period and a dash can be entered.
|
Press the +/- key and enter your METER ID#. Press ENTER to complete the entry.
|